Beer-Lambert Law Calculator
Apply the Beer-Lambert law, A = εlc, to turn absorbance into concentration or solve for any other term, with a flag for readings outside the linear range.
Calculator
Unitless. Keep it between 0.1 and 1.0, a safe working window.
Property of the substance at the measurement wavelength.
A standard cuvette is 1 cm.
- Transmittance %T = 100 × 10⁻ᴬ. Absorbance 1 means 10% of the light gets through.
- 22.39 %
- Light absorbed
- 77.61 %
Working, with your numbers
- c = A / (e × l)
- = 0.65 / (43,000 × 1)
- = 1.5116 × 10⁻⁵ mol/L
Values are converted into the units the equation is worked in before the arithmetic.
Use this answer in
Cite this tool
Citing this tool
Last updated . Add the date you accessed it as well, which a citation of a page that can change asks for. If a specific result matters, cite the permalink from the tool’s share row instead of this page: it reproduces the exact parameters.
On the syllabus: AP Chemistry, Unit 3: Properties of Substances and Mixtures.
The equation
Bouguer (1729), Lambert (1760) and Beer (1852), as reviewed by Mayerhöfer, Pahlow and Popp (2020)
The equation as an image to reuse
What is the Beer-Lambert law formula?
The Beer-Lambert law formula is A = εlc, where A is unitless optical absorbance, ε is molar absorptivity or extinction coefficient in M⁻¹cm⁻¹, l is sample cuvette path length in centimeters (cm), and c is solution concentration in moles per liter (M or mol/L). Rearranging solves for concentration: c = A / (εl).
What A = εlc is saying
In spectrophotometry, absorbance scales with how strongly
a species absorbs light, how far the light travels through the sample, and
how much of the species is present. In A = εlc the molar
extinction coefficient ε carries units of
M⁻¹cm⁻¹, the path length l is in centimeters,
and the concentration c is in mol/L. Those units cancel
exactly, so absorbance is unitless.
Because the relation is linear in concentration, a single measurement on a
known standard calibrates the instrument for that species. Rearranged for
the usual unknown, c = A / (εl).
Worked example
A protein with ε = 43000 M⁻¹cm⁻¹ is read at 15 µM in a standard 1 cm cuvette.
- Convert the concentration:
15 µM = 1.5e-5 M. A = ε l c = 43000 × 1 × 1.5e-5A = 0.645- Working backwards:
c = 0.645 / (43000 × 1) = 1.5e-5 M.
Absorbance relates to transmittance as A = −log10(T). So
A = 1 means 10% of the light passes through,
A = 2 means 1%, and A = 0.301 means 50%.
Staying inside the linear range
The law holds over a limited window of absorbance. The usable upper limit depends on the instrument, chiefly its stray light, and 0.1 to 1.0 is a safe working window: Qiagen gives that range for accurate readings at 260 nm. At 1.5 only about 3% of the incident light reaches the detector, so stray light leaking through the optics and electronic noise become comparable to the real signal. Measured absorbance then falls below the true value and the response curves away from a straight line.
Below about 0.1 the opposite problem applies: the difference between sample and blank becomes small, and small baseline drifts translate into large relative errors in concentration. The fix at the high end is dilution by a known factor, then multiplying the result back. At the low end, use a longer path length or concentrate the sample.
Common mistakes
- Reading outside the linear range. An absorbance of 2.8 is not four times more informative than 0.7; it is unreliable. Dilute into the working window rather than trusting the number.
- Skipping the blank. The instrument must be zeroed against the solvent and buffer in the same cuvette, otherwise the solvent’s own absorbance and the cuvette walls are counted as signal.
- Using ε from the wrong wavelength. Extinction coefficients are wavelength-specific and often pH- or solvent-specific. An ε quoted at 280 nm does not apply to a reading taken at 260 nm.
- Assuming a 1 cm path length. That is true of standard cuvettes but not of microplates, where the path length is set by the fill volume and well geometry. Measure or calculate it before applying the formula.
Sources
The 0.1 to 1.0 window: Qiagen,
DNeasy
Blood & Tissue Handbook (June 2023), appendix A, “Absorbance readings at 260 nm should fall
between 0.1 and 1.0 to be accurate.” The history of the law, from Bouguer and Lambert to Beer’s
1852 paper: Thomas Mayerhöfer, Susanne Pahlow and Jürgen Popp,
ChemPhysChem 21:2029 (2020). The
transmitted fractions follow from A = −log10(T): 10−1.5 is 3.2 percent.
Converting units first? Use the concentration and length conversion tables.
How to calculate concentration from absorbance
- Write down the formula:
c = A / (ε · l) -
Gather the values it needs. The calculator above starts with these:
- Absorbance: A = 0.65 AU
- Molar extinction coefficient: ε = 43,000 M⁻¹cm⁻¹
- Path length: l = 1 cm
-
Substitute them and work through the arithmetic:
- c = A / (e × l)
- = 0.65 / (43,000 × 1)
- = 1.5116 × 10⁻⁵ mol/L
-
To find one of the other values instead, rearrange the formula for it:
- Absorbance:
A = ε · l · c - Molar extinction coefficient:
ε = A / (l · c) - Path length:
l = A / (ε · c)
- Absorbance:
More worked examples
Each one runs through the calculator above, so the arithmetic here is the arithmetic it does.
What NADH concentration gives an absorbance of 0.311 at 340 nm in a 1 cm cuvette?
- c = A / (e × l)
- = 0.311 / (6220 × 1)
- = 5 × 10⁻⁵ mol/L
50 µM, using 6220 M⁻¹cm⁻¹, the molar absorptivity of NADH at 340 nm that dehydrogenase assays rely on. NAD⁺ barely absorbs at that wavelength, so the reading follows the reduced form alone. The same figure is often printed as 6.22 per millimolar, and mixing the two scales puts the answer out by a thousand.
What is the molar absorptivity if a 50 µM solution reads 0.6 in a 1 cm cuvette?
- e = A / (l × c)
- = 0.6 / (1 × 5 × 10⁻⁵)
- = 12,000 /M/cm
12,000 M⁻¹cm⁻¹, once the concentration is in moles per litre. Left as 50, it comes out as 0.012, a million times too small, with nothing about it that looks wrong. A single standard gives one point; reading several concentrations and taking the slope is more reliable, because a constant error in the blank drops out of a slope.
Common questions
Why should absorbance stay below about 1.0?
The usable upper limit depends on the instrument, chiefly its stray light, and 0.1 to 1.0 is a safe working window: Qiagen gives that range for accurate readings at 260 nm. At A = 1.5 only about 3 percent of the light reaches the detector, so stray light and detector noise become a large share of the signal and the response stops being linear. Dilute the sample and multiply back up rather than trusting a high reading.
How do absorbance and transmittance relate?
A = −log₁₀(T), so absorbance 1 means 10 percent transmitted, absorbance 2 means 1 percent, and absorbance 0 means fully transparent. The relationship is logarithmic, which is why small absorbance changes at the high end represent very large changes in transmitted light.