2:1 AV block on an EKG

A 6-second lead II strip of 2:1 AV block on ECG (EKG) paper, drawn by the ECG rhythm simulator’s model at 75 P waves per minute, with the guideline criteria and why the strip cannot say which type of second-degree block it is.

Lead II, 6 seconds at 25 mm/s and 10 mm/mV: each small square is 0.04 s and 0.1 mV, each large square 0.2 s and 0.5 mV, and the step at the left is the 1 mV calibration pulse. Filled triangles mark P waves that reached the ventricles, open triangles those that did not.

Is 2:1 AV block Mobitz type 1 or type 2?

Neither can be called from the strip alone. With every second P wave dropped there are never two conducted PR intervals in a row, so the PR cannot be seen to lengthen or to stay constant. The 2018 ACC/AHA/HRS guideline defines 2:1 block on its own and leaves it out of Mobitz type 2.

Defining criteria

Rates and intervals on this strip

Rates and intervals of the 2:1 AV block strip above, from the simulator’s model
Ventricular rate 37.5 bpm
Atrial rate 75 per minute
P waves conducted 1 of every 2
PR interval 160 ms
QRS duration 80 ms
QT interval 506 ms
QTc (Bazett) 400 ms
RR interval 1600 ms
Regularity Regular
Six-second count 4 QRS × 10 = 40 bpm

Measuring the strip

The ventricular rate is half the atrial rate, 75 / 2 = 37.5 beats per minute, and every RR interval is two PP intervals, 2 × 800 = 1600 ms. Every conducted beat has the same PR, 160 ms, so the ventricular rhythm is regular.

The dropped P wave starts 134 ms after the T wave ends, alone on the baseline between two beats. With a faster atrial rate and a longer PR it moves into the T wave, where the strip can pass for a slow sinus rhythm until the P waves are marched out with calipers.

A narrow QRS points to a block in the AV node and a wide one to the conduction system below it, which is how the simulator’s ladder decides where to draw the block. Telling them apart with certainty needs more than this strip, such as a longer recording that catches two conducted beats in a row.

Settings the simulator allows for this rhythm

Each rhythm limits its settings to the values its definition allows, so the simulator can never draw it as something else.

Settings the ECG simulator allows for 2:1 AV block, and the value the strip above uses
Setting Allowed This strip
Sinus rate 40 to 99 bpm 75 bpm
PR interval 120 to 280 ms 160 ms
QRS duration 60 to 160 ms 80 ms

Telling it apart

Sources

  1. Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS Guideline on the Evaluation and Management of Patients With Bradycardia and Cardiac Conduction Delay. Circulation. 2019;140(8):e382-e482. doi:10.1161/CIR.0000000000000628
  2. Kligfield P, Gettes LS, Bailey JJ, et al. Recommendations for the Standardization and Interpretation of the Electrocardiogram, part I. Circulation. 2007;115:1306-1324. doi:10.1161/CIRCULATIONAHA.106.180200
  3. McSharry PE, Clifford GD, Tarassenko L, Smith LA. A dynamical model for generating synthetic electrocardiogram signals. IEEE Transactions on Biomedical Engineering. 2003;50(3):289-294. doi:10.1109/TBME.2003.808805
  4. Bazett HC. An analysis of the time-relations of electrocardiograms. Heart. 1920;7:353-370.