A 6-second lead II strip of atrial fibrillation on ECG (EKG) paper at a mean ventricular rate of 100, drawn by the ECG rhythm simulator’s model, with the ESC criteria and every RR interval measured.

Lead II, 6 seconds at 25 mm/s and 10 mm/mV: each small square is 0.04 s and 0.1 mV, each large square 0.2 s and 0.5 mV, and the step at the left is the 1 mV calibration pulse.

What does atrial fibrillation look like on an EKG?

No P waves, only an irregular wavering baseline, and RR intervals that are irregularly irregular, with no pattern to the spacing of the QRS complexes. The QRS is usually narrow. The ventricular rate depends on how many fibrillatory impulses the AV node lets through, so it can be slow, normal or fast.

Defining criteria

Rates and intervals on this strip

Rates and intervals of the atrial fibrillation strip above, from the simulator’s model
Ventricular rate 100 bpm (mean over the minute)
Atrial rate No P waves
PR interval None, no P waves
QRS duration 80 ms
QT interval 310 ms
QTc (Bazett) 400 ms
RR interval over the minute 450 to 1044 ms
RR interval on this strip 450 to 718 ms
Regularity Irregularly irregular
Six-second count 11 QRS × 10 = 110 bpm

Measuring the strip

The mean RR interval is 60 / 100 = 600 ms. The model makes the mean wait for the next impulse 60 / 400 = 150 ms, so the AV node’s refractory period is 600 − 150 = 450 ms and no RR interval is shorter. Over the whole minute they run from 450 to 1044 ms.

On this strip alone the RR intervals run from 450 to 718 ms. Six seconds hold 11 QRS complexes, so the six-second count gives 110 for a rate that averages 100 over the minute. In an irregular rhythm no single RR interval stands for the rate, and a longer strip or a full minute gives a better average.

The baseline is the sum of four waves between 350 and 540 per minute and never strays more than 0.06 mV from zero, under one small square. On a real tracing fibrillatory waves can be coarser, or so fine that the baseline looks flat, so the RR intervals are the surer sign.

Settings the simulator allows for this rhythm

Each rhythm limits its settings to the values its definition allows, so the simulator can never draw it as something else.

Settings the ECG simulator allows for atrial fibrillation, and the value the strip above uses
Setting Allowed This strip
Mean ventricular rate 40 to 180 bpm 100 bpm
QRS duration 60 to 160 ms 80 ms

Telling it apart

Sources

  1. Hindricks G, Potpara T, Dagres N, et al. 2020 ESC Guidelines for the diagnosis and management of atrial fibrillation. European Heart Journal. 2021;42(5):373-498. doi:10.1093/eurheartj/ehaa612
  2. Kligfield P, Gettes LS, Bailey JJ, et al. Recommendations for the Standardization and Interpretation of the Electrocardiogram, part I. Circulation. 2007;115:1306-1324. doi:10.1161/CIRCULATIONAHA.106.180200
  3. McSharry PE, Clifford GD, Tarassenko L, Smith LA. A dynamical model for generating synthetic electrocardiogram signals. IEEE Transactions on Biomedical Engineering. 2003;50(3):289-294. doi:10.1109/TBME.2003.808805
  4. Bazett HC. An analysis of the time-relations of electrocardiograms. Heart. 1920;7:353-370.

The values are facts and free to use; this page’s selection and presentation are © 2026 ScienceQuest.