Mobitz type 2 AV block on an EKG
A 6-second lead II strip of second-degree AV block, Mobitz type 2, on ECG (EKG) paper with 4:3 conduction, drawn by the ECG rhythm simulator’s model, with the guideline criteria and the pause across the dropped beat.
What does Mobitz type 2 look like on an EKG?
Every conducted beat has the same PR interval, before and after a P wave that suddenly fails to reach the ventricles, with no lengthening of the PR to warn of it. The block is usually below the AV node, in the His-Purkinje system, often with a wide QRS, and it can progress to complete heart block.
Defining criteria
- Second-degree AV block: P waves at a constant rate below 100 per minute, with AV conduction present but not 1:1. (Kusumoto et al. 2018)
- Mobitz type 2: a single P wave periodically not conducted, with constant PR intervals before and after it, 2:1 block excluded. (Kusumoto et al. 2018)
- The simulator allows P wave rates of 40 to 99 and groups of 3:2, 4:3, 5:4 or 6:5. The PR may be 120 to 280 ms. (simulator model)
Rates and intervals on this strip
| Ventricular rate | 56.25 bpm |
|---|---|
| Atrial rate | 75 per minute |
| P waves conducted | 3 of every 4 |
| PR interval | 160 ms |
| QRS duration | 80 ms |
| QT interval | 413 ms |
| QTc (Bazett) | 400 ms |
| RR interval | 800 to 1600 ms |
| Regularity | Regularly irregular |
| Six-second count | 6 QRS × 10 = 60 bpm |
Measuring the group
The PR is 160 ms in every conducted beat, so the conducted beats come every 800 ms, the PP interval. The dropped P wave leaves a gap of exactly 2 × 800 = 1600 ms, and 3 QRS complexes for every 4 P waves give 75 × 3 / 4 = 56.25 beats per minute.
The 2018 guideline recommends a permanent pacemaker for acquired Mobitz type 2 with no reversible or physiologic cause, whether or not it causes symptoms. Widen the QRS to 120 ms or more in the simulator to see the wide QRS it most often comes with.
Settings the simulator allows for this rhythm
Each rhythm limits its settings to the values its definition allows, so the simulator can never draw it as something else.
| Setting | Allowed | This strip |
|---|---|---|
| Sinus rate | 40 to 99 bpm | 75 bpm |
| PR interval | 120 to 280 ms | 160 ms |
| QRS duration | 60 to 160 ms | 80 ms |
| Conduction ratio | 3:2, 4:3, 5:4 or 6:5 | 4:3 |
Telling it apart
- In Mobitz type 1 the PR lengthens before the dropped P wave and the first PR after it is the shortest.
- 2:1 AV block has no two conducted PR intervals in a row to compare, so it is excluded from Mobitz type 2.
- When no P wave reaches the ventricles at all, the rhythm is complete heart block.
Sources
- Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS Guideline on the Evaluation and Management of Patients With Bradycardia and Cardiac Conduction Delay. Circulation. 2019;140(8):e382-e482. doi:10.1161/CIR.0000000000000628
- Kligfield P, Gettes LS, Bailey JJ, et al. Recommendations for the Standardization and Interpretation of the Electrocardiogram, part I. Circulation. 2007;115:1306-1324. doi:10.1161/CIRCULATIONAHA.106.180200
- McSharry PE, Clifford GD, Tarassenko L, Smith LA. A dynamical model for generating synthetic electrocardiogram signals. IEEE Transactions on Biomedical Engineering. 2003;50(3):289-294. doi:10.1109/TBME.2003.808805
- Bazett HC. An analysis of the time-relations of electrocardiograms. Heart. 1920;7:353-370.
The values are facts and free to use; this page’s selection and presentation are © 2026 ScienceQuest.